Automated Detection of Near Falls: Algorithm Development and Preliminary Results

DSpace/Manakin Repository

Automated Detection of Near Falls: Algorithm Development and Preliminary Results

Citable link to this page

. . . . . .

Title: Automated Detection of Near Falls: Algorithm Development and Preliminary Results
Author: Weiss, Aner; Shimkin, Ilan; Giladi, Nir; Hausdorff, Jeffrey M.

Note: Order does not necessarily reflect citation order of authors.

Citation: Weiss, Aner, Ilan Shimkin, Nir Giladi, and Jeffrey M. Hausdorff. 2010. Automated detection of near falls: Algorithm development and preliminary results. BMC Research Notes 3: 62.
Full Text & Related Files:
Abstract: Background: Falls are a major source of morbidity and mortality among older adults. Unfortunately, self-report is, to a large degree, the gold-standard method for characterizing and quantifying fall frequency. A number of studies have demonstrated that near falls predict falls and that near falls may occur more frequently than falls. These studies suggest that near falls might be an appropriate fall risk measure. However, to date, such investigations have also relied on self-report. The purpose of the present study was to develop a method for automatic detection of near falls, potentially a sensitive, objective marker of fall risk and to demonstrate the ability to detect near falls using this approach. Findings: 15 healthy subjects wore a tri-axial accelerometer on the pelvis as they walked on a treadmill under different conditions. Near falls were induced by placing obstacles on the treadmill and were defined using observational analysis. Acceleration-derived parameters were examined as potential indicators of near falls, alone and in various combinations. 21 near falls were observed and compared to 668 "non-near falls" segments, consisting of normal and abnormal (but not near falls) gait. The best single method was based on the maximum peak-to-peak vertical acceleration derivative, with detection rates better than 85% sensitivity and specificity. Conclusions: These findings suggest that tri-axial accelerometers may be used to successfully distinguish near falls from other gait patterns observed in the gait laboratory and may have the potential for improving the objective evaluation of fall risk, perhaps both in the lab and in at home-settings.
Published Version: doi:10.1186/1756-0500-3-62
Other Sources: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC2845599/pdf/
Terms of Use: This article is made available under the terms and conditions applicable to Other Posted Material, as set forth at http://nrs.harvard.edu/urn-3:HUL.InstRepos:dash.current.terms-of-use#LAA
Citable link to this page: http://nrs.harvard.edu/urn-3:HUL.InstRepos:5310905

Show full Dublin Core record

This item appears in the following Collection(s)

 
 

Search DASH


Advanced Search
 
 

Submitters