Publication:
X-ray monochromator system for use with synchrotron radiation sources

Thumbnail Image

Date

1981

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

AIP Publishing
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Golovchenko, J. A., Levesque, R. A., Cowan, P. L. 1981. X-Ray Monochromator System for Use with Synchrotron Radiation Sources. Review of Scientific Instruments. doi:10.1063/1.1136631.

Research Data

Abstract

A double crystal x‐ray monochromator system has been developed for use on synchrotron radiationsources. The system consists of two separate Bragg reflecting crystals constrained by a mechanical linkage system enabling it to tune continuously in x‐ray energy, always maintaining an exit beam of constant offset and direction relative to the incident beam. Accurate parallelism between the two crystals is maintained by a piezoelectrically‐controlled analog feedback circuit.

Description

Keywords

monochromators, radioactive sources, synchrotron radiation, mechanical energy