Liquid crystal phantom for validation of microscopic diffusion anisotropy measurements on clinical MRI systems

Thumbnail Image

Date

2017

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

John Wiley and Sons Inc.
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Nilsson, Markus, Johan Larsson, Dan Lundberg, Filip Szczepankiewicz, Thomas Witzel, Carl‐Fredrik Westin, Karin Bryskhe, and Daniel Topgaard. 2017. “Liquid crystal phantom for validation of microscopic diffusion anisotropy measurements on clinical MRI systems.” Magnetic Resonance in Medicine 79 (3): 1817-1828. doi:10.1002/mrm.26814. http://dx.doi.org/10.1002/mrm.26814.

Research Data

Abstract

Purpose To develop a phantom for validating MRI pulse sequences and data processing methods to quantify microscopic diffusion anisotropy in the human brain. Methods: Using a liquid crystal consisting of water, detergent, and hydrocarbon, we designed a 0.5‐L spherical phantom showing the theoretically highest possible degree of microscopic anisotropy. Data were acquired on the Connectome scanner using echo‐planar imaging signal readout and diffusion encoding with axisymmetric b‐tensors of varying magnitude, anisotropy, and orientation. The mean diffusivity, fractional anisotropy (FA), and microscopic FA (µFA) parameters were estimated. Results: The phantom was observed to have values of mean diffusivity similar to brain tissue, and relaxation times compatible with echo‐planar imaging echo times on the order of 100 ms. The estimated values of µFA were at the theoretical maximum of 1.0, whereas the values of FA spanned the interval from 0.0 to 0.8 as a result of varying orientational order of the anisotropic domains within each voxel. Conclusions: The proposed phantom can be manufactured by mixing three widely available chemicals in volumes comparable to a human head. The acquired data are in excellent agreement with theoretical predictions, showing that the phantom is ideal for validating methods for measuring microscopic diffusion anisotropy on clinical MRI systems. Magn Reson Med 79:1817–1828, 2018. © 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution‐NonCommercial License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited and is not used for commercial purposes.

Description

Keywords

Full Paper, dMRI, QTE, gamma model, powder averaging, microstructure