MRI-Based Multiscale Model for Electromagnetic Analysis in the Human Head with Implanted DBS

DSpace/Manakin Repository

MRI-Based Multiscale Model for Electromagnetic Analysis in the Human Head with Implanted DBS

Citable link to this page

 

 
Title: MRI-Based Multiscale Model for Electromagnetic Analysis in the Human Head with Implanted DBS
Author: Iacono, Maria Ida; Makris, Nikos; Mainardi, Luca; Angelone, Leonardo M.; Bonmassar, Giorgio

Note: Order does not necessarily reflect citation order of authors.

Citation: Iacono, Maria Ida, Nikos Makris, Luca Mainardi, Leonardo M. Angelone, and Giorgio Bonmassar. 2013. “MRI-Based Multiscale Model for Electromagnetic Analysis in the Human Head with Implanted DBS.” Computational and Mathematical Methods in Medicine 2013 (1): 694171. doi:10.1155/2013/694171. http://dx.doi.org/10.1155/2013/694171.
Full Text & Related Files:
Abstract: Deep brain stimulation (DBS) is an established procedure for the treatment of movement and affective disorders. Patients with DBS may benefit from magnetic resonance imaging (MRI) to evaluate injuries or comorbidities. However, the MRI radio-frequency (RF) energy may cause excessive tissue heating particularly near the electrode. This paper studies how the accuracy of numerical modeling of the RF field inside a DBS patient varies with spatial resolution and corresponding anatomical detail of the volume surrounding the electrodes. A multiscale model (MS) was created by an atlas-based segmentation using a 1 mm3 head model (mRes) refined in the basal ganglia by a 200 μm2 ex-vivo dataset. Four DBS electrodes targeting the left globus pallidus internus were modeled. Electromagnetic simulations at 128 MHz showed that the peak of the electric field of the MS doubled (18.7 kV/m versus 9.33 kV/m) and shifted 6.4 mm compared to the mRes model. Additionally, the MS had a sixfold increase over the mRes model in peak-specific absorption rate (SAR of 43.9 kW/kg versus 7 kW/kg). The results suggest that submillimetric resolution and improved anatomical detail in the model may increase the accuracy of computed electric field and local SAR around the tip of the implant.
Published Version: doi:10.1155/2013/694171
Other Sources: http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3727211/pdf/
Terms of Use: This article is made available under the terms and conditions applicable to Other Posted Material, as set forth at http://nrs.harvard.edu/urn-3:HUL.InstRepos:dash.current.terms-of-use#LAA
Citable link to this page: http://nrs.harvard.edu/urn-3:HUL.InstRepos:11855908