Myocardial Defect Detection Using PET-CT: Phantom Studies
View/ Open
Published Version
https://doi.org/10.1371/journal.pone.0088200Metadata
Show full item recordCitation
Mananga, Eugene S., Georges El Fakhri, Joshua Schaefferkoetter, Ali A. Bonab, and Jinsong Ouyang. 2014. “Myocardial Defect Detection Using PET-CT: Phantom Studies.” PLoS ONE 9 (2): e88200. doi:10.1371/journal.pone.0088200. http://dx.doi.org/10.1371/journal.pone.0088200.Abstract
It is expected that both noise and activity distribution can have impact on the detectability of a myocardial defect in a cardiac PET study. In this work, we performed phantom studies to investigate the detectability of a defect in the myocardium for different noise levels and activity distributions. We evaluated the performance of three reconstruction schemes: Filtered Back-Projection (FBP), Ordinary Poisson Ordered Subset Expectation Maximization (OP–OSEM), and Point Spread Function corrected OSEM (PSF–OSEM). We used the Channelized Hotelling Observer (CHO) for the task of myocardial defect detection. We found that the detectability of a myocardial defect is almost entirely dependent on the noise level and the contrast between the defect and its surroundings.Other Sources
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3914931/pdf/Terms of Use
This article is made available under the terms and conditions applicable to Other Posted Material, as set forth at http://nrs.harvard.edu/urn-3:HUL.InstRepos:dash.current.terms-of-use#LAACitable link to this page