Yamaki, TakahikoKawasaki, MasanoriRaffel, Owen ChristopherIshihara, YoshiyukiOkubo, MunenoriKubota, TomokiHattori, ArihiroNishigaki, KazuhikoTakemura, GenzouFujiwara, HisayoshiMinatoguchi, ShinyaJang, Ik-Kyung2013-04-222012Yamaki, Takahiko, Masanori Kawasaki, Ik-Kyung Jang, Owen Christopher Raffel, Yoshiyuki Ishihara, Munenori Okubo, Tomoki Kubota, et al. 2012. Comparison between integrated backscatter intravascular ultrasound and 64-slice multi-detector row computed tomography for tissue characterization and volumetric assessment of coronary plaques. Cardiovascular Ultrasound 10:33.1476-7120http://nrs.harvard.edu/urn-3:HUL.InstRepos:10579214Background: The purpose of this study was to determine the cut-off values of Hounsfield units (HU) for the discrimination of plaque components and to evaluate the feasibility of measurement of the volume of plaque components using multi-detector row computed tomography (MDCT). Methods: Coronary lesions (125 lesions in 125 patients) were visualized by both integrated backscatter intravascular ultrasound (IB-IVUS) and 64-slice MDCT at the same site. The IB values were used as a gold standard to determine the cut off values of HU for the discrimination of plaque components. Results: Plaques were classified as lipid pool (n =50), fibrosis (n =65) or calcification (n =35) by IB-IVUS. The HU of lipid pool, fibrosis and calcification were 18 ± 18 HU (−19 to 58 HU), 95 ± 24 HU (46 to 154 HU) and 378 ± 99 HU (188 to 605 HU), respectively. Using receiver operating characteristic curve analysis, a threshold of 50 HU was the optimal cutoff values to discriminate lipid pool from fibrosis. Lipid volume measured by MDCT was correlated with that measured by IB-IVUS (r =0.66, p <0.001), whereas fibrous volume was not (r =0.21, p =0.059). Conclusion: Lipid volume measured by MDCT was moderately correlated with that measured by IB-IVUS. MDCT may be useful for volumetric assessment of the lipid volume of coronary plaques, whereas the assessment of fibrosis volume was unstable.en-USComputed tomographyIntegrated backscatterIntravascular ultrasoundCoronary plaqueComparison between Integrated Backscatter Intravascular Ultrasound and 64-Slice Multi-Detector Row Computed Tomography for Tissue Characterization and Volumetric Assessment of Coronary PlaquesJournal Article2013-04-2210.1186/1476-7120-10-33