Publication:

Comparison of Rates of Safety Issues and Reporting of Trial Outcomes for Medical Devices Approved in the European Union and United States: Cohort Study

Loading...
Thumbnail Image

Date

2016-06-28

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

BMJ
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Hwang, Thomas J, Elisaveta Sokolov, Jessica M Franklin, and Aaron S Kesselheim. 2016. “Comparison of Rates of Safety Issues and Reporting of Trial Outcomes for Medical Devices Approved in the European Union and United States: Cohort Study.” BMJ, June, i3323. https://doi.org/10.1136/bmj.i3323.

Abstract

OBJECTIVE To evaluate safety alerts and recalls, publication of key trial outcomes, and subsequent US approval of high profile medical devices introduced in the European Union. DESIGN Cohort study. SETTING Novel cardiovascular, orthopedic, and neurologic devices approved in the EU through Conformite Europeenne marking between 2005 and 2010. DATA SOURCES Public and commercial databases searched up to January 2016 for press releases and announcements of approvals; public Food and Drug Administration and European regulatory authority databases for US approvals and safety alerts and recalls; and Medline, Embase, and Web of Science for peer reviewed publications. MAIN OUTCOME MEASURES We categorized the novelty of the devices in the study sample as a "major innovation" or an "other change," and extracted descriptive data about the devices and information on any safety alerts and withdrawals. Linear regression models examined factors associated with differential EU and US approvals. Cox proportional hazards regression models were used to evaluate factors associated with safety alerts and recalls and the publication of trial outcomes for devices categorized as major innovations. Models controlled for time, therapeutic category, regulatory pathway, size of sponsoring company, and indicator variables for devices approved first in the EU and devices approved only in the EU.RESULTS67% (206/309) of devices identified were approved in both the US and the EU, of which 63% (129/206) were approved first in the EU. The unadjusted rate of safety alerts and recalls for devices approved first in the EU was 27% (62/232) compared with 14% (11/77) for devices approved first in the US. The adjusted hazard ratio for safety alerts and recalls was 2.9 (95% confidence interval 1.4 to 6.2) for devices approved first in the EU. The results of pivotal trials were published for 49% (37/75) of devices categorized as major innovations, with an overall publication rate of 37% five years after approval. CONCLUSIONS Devices approved first in the EU are associated with an increased risk of post-marketing safety alerts and recalls. Poor trial publication rates mean that patients and clinicians need greater regulatory transparency to make informed decisions about treatment.

Description

Other Available Sources

Research Data

Keywords

Terms of Use

This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service

Endorsement

Review

Supplemented By

Related Stories