Publication:
A Low-cost, MR-compatible Olfactometer

Thumbnail Image

Date

2006

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

Psychonomic Society
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Lowen, Steven B., and Scott E. Lukas. 2006. A low-cost, MR-compatible olfactometer. Behavior Research Methods 38(2): 307-313.

Research Data

Abstract

We present a design for an olfactometer, suitable for fMRI experiments, that can be constructed at extremely low cost. The olfactometer presents odors directly to the nose via a nasal cannula at unobtrusively low flow velocities, with no large assemblies required on or near the subject's face. The olfactometer can be controlled manually, or by computer via a serial interface. A validation study verified that the olfactometer reliably presents odors to test subjects. Errors and response latency times decreased with increased flow rate in an orderly manner, as expected.