A Systematic Comparison of Motion Artifact Correction Techniques for Functional Near-Infrared Spectroscopy

Thumbnail Image

Open/View Files

Date

2012

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

Frontiers Media S.A.
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Cooper, Robert J., Juliette Selb, Louis Gagnon, Dorte Phillip, Henrik W. Schytz, Helle K. Iversen, Messoud Ashina, and David A. Boas. 2012. A systematic comparison of motion artifact correction techniques for functional near-infrared spectroscopy. Frontiers in Neuroscience 6:147.

Research Data

Abstract

Near-infrared spectroscopy (NIRS) is susceptible to signal artifacts caused by relative motion between NIRS optical fibers and the scalp. These artifacts can be very damaging to the utility of functional NIRS, particularly in challenging subject groups where motion can be unavoidable. A number of approaches to the removal of motion artifacts from NIRS data have been suggested. In this paper we systematically compare the utility of a variety of published NIRS motion correction techniques using a simulated functional activation signal added to 20 real NIRS datasets which contain motion artifacts. Principle component analysis, spline interpolation, wavelet analysis, and Kalman filtering approaches are compared to one another and to standard approaches using the accuracy of the recovered, simulated hemodynamic response function (HRF). Each of the four motion correction techniques we tested yields a significant reduction in the mean-squared error (MSE) and significant increase in the contrast-to-noise ratio (CNR) of the recovered HRF when compared to no correction and compared to a process of rejecting motion-contaminated trials. Spline interpolation produces the largest average reduction in MSE (55%) while wavelet analysis produces the highest average increase in CNR (39%). On the basis of this analysis, we recommend the routine application of motion correction techniques (particularly spline interpolation or wavelet analysis) to minimize the impact of motion artifacts on functional NIRS data.

Description

Keywords

near-infrared spectroscopy, functional near-infrared spectroscopy, NIRS, motion artifact, hemodynamic response