Publication: Accuracy and Precision in Quantitative Fluorescence Microscopy
Open/View Files
Date
2009
Authors
Published Version
Journal Title
Journal ISSN
Volume Title
Publisher
The Rockefeller University Press
The Harvard community has made this article openly available. Please share how this access benefits you.
Citation
Waters, Jennifer C. 2009. Accuracy and precision in quantitative fluorescence microscopy. The Journal of Cell Biology 185(7): 1135-1148.
Research Data
Abstract
The light microscope has long been used to document the localization of fluorescent molecules in cell biology research. With advances in digital cameras and the discovery and development of genetically encoded fluorophores, there has been a huge increase in the use of fluorescence microscopy to quantify spatial and temporal measurements of fluorescent molecules in biological specimens. Whether simply comparing the relative intensities of two fluorescent specimens, or using advanced techniques like Förster resonance energy transfer (FRET) or fluorescence recovery after photobleaching (FRAP), quantitation of fluorescence requires a thorough understanding of the limitations of and proper use of the different components of the imaging system. Here, I focus on the parameters of digital image acquisition that affect the accuracy and precision of quantitative fluorescence microscopy measurements.