Publication:
Design of a Motion Compensated Tissue Resection Catheter for Beating Heart Cardiac Surgery

Thumbnail Image

Date

2011

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

ASME International
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Kesner, Samuel B., and Robert D. Howe. 2011. “Design of a Motion Compensated Tissue Resection Catheter for Beating Heart Cardiac Surgery.” Journal of Medical Devices 5 (2): 027523. doi:10.1115/1.3590649.

Research Data

Abstract

Cardiac catheters allow clinicians to minimally invasively interact with the beating heart without stopping the heart or opening the chest. However, the fast motion of the intracardiac structures makes it difficult to modify and repair the tissue in a controlled and safe manner. To enable surgical procedures on the inside of the beating heart, we have developed an ultrasound-guided catheter system that virtually freezes the heart by compensating for the fast cardiac motions. The device presented in this paper is a resection tool that allows the catheter system to cut moving tissue, a key surgical task required for many intracardiac procedures including valve and leaflet repair. The motion tracking system is demonstrated in vivo and the tissue resection tool is evaluated by resecting tissue mounted on a cardiac motion simulator. The motion compensated catheter is shown to greatly improve the resection cut quality on the moving tissue target while reducing the forces experienced by the tissue by almost 80%

Description

Other Available Sources

Keywords

Terms of Use

This article is made available under the terms and conditions applicable to Open Access Policy Articles (OAP), as set forth at Terms of Service

Endorsement

Review

Supplemented By

Referenced By

Related Stories