Publication:

An Apparatus for Performing Microtensile Tests at Elevated Temperatures inside a Scanning Electron Microscope

Loading...
Thumbnail Image

Date

2013

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Sim, Gi-Dong, Jun-Hyub Park, Michael D. Uchic, Paul A. Shade, Soon-Bok Lee, and Joost J. Vlassak. 2013. An apparatus for performing microtensile tests at elevated temperatures inside a scanning electron microscope. Acta Materialia 61, no. 19: 7500-7510.

Abstract

In this paper, we introduce an apparatus to perform microtensile tests at elevated temperatures inside a scanning electron microscope. The apparatus has a stroke of 250 μm with a displacement resolution of 10 nm and a load resolution of 9.7 μN. Measurements at elevated temperatures are performed through use of two silicon-based micromachined heaters that support the sample. Each heater consists of a tungsten heating element that also serves as a temperature gauge. To demonstrate the testing capabilities, tensile tests were performed on submicron Cu films at various temperatures up to 430 °C. Stress–strain curves show a significant decrease in yield strength and initial slope for the samples tested at elevated temperature, which we attribute to diffusion-facilitated grain boundary sliding and dislocation climb.

Description

Other Available Sources

Research Data

Keywords

In situ tension test, High-temperature deformation, Thin film, Grain boundary sliding, Creep