Publication: Influence of Image Acquisition on Radiation Dose and Image Quality: Full versus Narrow Phase Window Acquisition Using 320 MDCT
Open/View Files
Date
2013
Published Version
Journal Title
Journal ISSN
Volume Title
Publisher
Hindawi Publishing Corporation
The Harvard community has made this article openly available. Please share how this access benefits you.
Citation
Khosa, Faisal, Atif Khan, Khurram Nasir, Waqas Shuaib, Matthew Budoff, Ron Blankstein, and Melvin E. Clouse. 2013. “Influence of Image Acquisition on Radiation Dose and Image Quality: Full versus Narrow Phase Window Acquisition Using 320 MDCT.” The Scientific World Journal 2013 (1): 731590. doi:10.1155/2013/731590. http://dx.doi.org/10.1155/2013/731590.
Research Data
Abstract
Purpose. To compare radiation dose and image quality using predefined narrow phase window versus complete phase window with dose modulation during R-R using 320-row MDCTA. Methods:. 114 patients underwent coronary CTA study using 320-row MDCT scanner. 87 patients with mean age (61 + 13 years), mean BMI (29 + 6), and mean heart rate (HR) (58 + 7 bpm) were imaged at predefined 66–80% R-R interval and then reconstructed at 75% while 27 patients with mean age (63 + 16 years), mean BMI (28 + 5), and mean HR (57 + 7 bpm) were scanned throughout the complete R-R interval with tube current modulation. The effective dose (ED) was calculated from dose length product (DLP) and conversion k (0.014 mSv/mGy/cm). Image quality was assessed using a three-point ordinal scale (1 = excellent, 2 = good, and 3 = nondiagnostic). Results:. Both groups were statistically similar to each other with reference of HR (P = 0.59), BMI (P = 0.17), and tube current mAs (P = 0.68). The median radiation dose was significantly higher in those scanned with complete R-R phase window versus narrow phase window (P < 0.0001). Independently of patient and scan parameters, increased phase window was associated with higher radiation dose (P < 0.001). Image quality was better among those scanned with narrow phase window versus complete phase window (P < 0.0001). Conclusion:. Our study supports that good HR control and predefined narrow window acquisition result in lower radiation dose without compromising diagnostic image quality for coronary disease evaluation.
Description
Other Available Sources
Keywords
Terms of Use
This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service