Vella, Sarah J., Patrick Beattie, Rebecca Cademartiri, Anna Laromaine, Andres W. Martinez, Scott T. Phillips, Katherine A. Mirica, and George M. Whitesides. 2012. Measuring Markers of Liver Function Using a Micropatterned Paper Device Designed for Blood from a Fingerstick. Analytical Chemistry 84, no. 6: 2883–2891.

Abstract

This paper describes a paper-based microfluidic device that measures two enzymatic markers of liver function (alkaline phosphatase, ALP, and aspartate aminotransferase, AST) and total serum protein. A device consists of four components: (i) a top plastic sheet, (ii) a filter membrane, (iii) a patterned paper chip containing the reagents necessary for analysis, and (iv) a bottom plastic sheet. The device performs both the sample preparation (separating blood plasma from erythrocytes) and the assays; it also enables both qualitative and quantitative analysis of data. The data obtained from the paper-microfluidic devices show standard deviations in calibration runs and “spiked” standards that are acceptable for routine clinical use. This device illustrates a type of test useable for a range of assays in resource-poor settings.

Description

Research Data

Keywords

Terms of Use

This article is made available under the terms and conditions applicable to Open Access Policy Articles (OAP), as set forth at Terms of Service

Endorsement

Review

Supplemented By

Related Stories