Evaluation of the NucliSens EasyQ v2.0 Assay in Comparison with the Roche Amplicor v1.5 and the Roche CAP/CTM HIV-1 Test v2.0 in Quantification of C-Clade HIV-1 in Plasma

Thumbnail Image

Date

2014

Journal Title

Journal ISSN

Volume Title

Publisher

Public Library of Science
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Muenchhoff, M., S. Madurai, A. J. Hempenstall, E. Adland, A. Carlqvist, A. Moonsamy, M. Jaggernath, et al. 2014. “Evaluation of the NucliSens EasyQ v2.0 Assay in Comparison with the Roche Amplicor v1.5 and the Roche CAP/CTM HIV-1 Test v2.0 in Quantification of C-Clade HIV-1 in Plasma.” PLoS ONE 9 (8): e103983. doi:10.1371/journal.pone.0103983. http://dx.doi.org/10.1371/journal.pone.0103983.

Research Data

Abstract

Human immunodeficiency virus type 1 (HIV-1) genetic diversity poses a challenge to reliable viral load monitoring. Discrepancies between different testing platforms have been observed, especially for non-clade-B virus. Therefore we compare, in antiretroviral therapy (ART)-naïve South African subjects predominantly infected with HIV-1 clade-C, three commercially available assays: the COBAS AmpliPrep/COBAS TaqMan HIV-1 Test version 2.0 by Roche (CAP/CTM v2.0), the BioMérieux NucliSens Version 2.0 Easy Q/Easy Mag (NucliSens v2.0) and the Roche COBAS Amplicor HIV-1 Monitor Test Version 1.5 (Amplicor v1.5). Strong linear correlation was observed and Bland-Altman analyses showed overall good agreement between the assays with mean viral load differences of 0.078 log cp/ml (NucliSens v2.0 – Amplicor v1.5), 0.260 log cp/ml (CAP/CTM v2.0 – Amplicor v1.5) and 0.164 log cp/ml (CAP/CTM v2.0 – NucliSens v2.0), indicating lower mean viral load results for the Amplicor v1.5 and higher mean readings for the CAP/CTM v2.0. Consistent with observations following previous comparisons of CAP/CTM v2.0 versus Amplicor v1.5, the CAP/CTM v2.0 assay detected low-level viremia (median 65 cp/ml) in more than one-third of those in whom viremia had been undetectable (<20 cp/ml) in assays using the NucliSens platform. These levels of viremia are of uncertain clinical significance but may be of importance in early detection of ART resistance in those on treatment. Overall the three assays showed good comparability of results but with consistent, albeit relatively small, discrepancies for HIV-1 clade-C samples, especially in the low-viremic range that should be taken into account when interpreting viral load data.

Description

Keywords

Medicine and health sciences, Infectious diseases, Viral diseases, HIV infections

Terms of Use

This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service

Endorsement

Review

Supplemented By

Referenced By

Related Stories