Publication:

Real-World Safety and Effectiveness Outcomes of a Zotarolimus-Eluting Stent: Final 3-Year Report of the RESOLUTE International Study

Loading...
Thumbnail Image

Date

2013

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

BlackWell Publishing Ltd
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

BELARDI, JORGE A, PETR WIDIMSKÝ, FRANZ-JOSEF NEUMANN, LAURA MAURI, and MARIANO ALBERTAL. 2013. “Real-World Safety and Effectiveness Outcomes of a Zotarolimus-Eluting Stent: Final 3-Year Report of the RESOLUTE International Study.” Journal of Interventional Cardiology 26 (5): 515-523. doi:10.1111/joic.12051. http://dx.doi.org/10.1111/joic.12051.

Abstract

Objectives: We evaluated the safety and effectiveness of the Resolute™ zotarolimus-eluting stent (R-ZES) in real-world clinical practice through 3 years. Background: A randomized comparison of the R-ZES and the XIENCE V™ everolimus-eluting stent showed no difference in any outcomes through 3-year follow-up in high-volume academic centers. RESOLUTE International is a confirmatory trial designed to evaluate the R-ZES in real-world clinical practice. Methods: RESOLUTE International is a single arm, observational trial that enrolled 2,349 patients from 88 centers with only a few inclusion and exclusion criteria. The primary end-point was the composite of cardiac death and target vessel myocardial infarction (TV-MI) at 1 year. Secondary end-points include target lesion failure (TLF), target vessel revascularization (TVR), and their components, and stent thrombosis (ST). Results: At 3 years 97.2% of patients completed clinical follow-up. The mean age was 63.4 ± 11.2 years, 77.8% were male, and 30.4% had diabetes. The average number of stents per patient was 1.6 ± 1.0; and mean stent length was 30.9 ± 20.5 mm. Dual antiplatelet therapy was used in 91.1% of patients at 1 year, 43.0% at 2 years, and 34.6% at 3 years. Cardiac death and TV-MI occurred in 161 patients (7.0%). There were 6 (0.3%) very late ST events for a total ST rate of 1.1% through 3 years. The rates of clinically driven target lesion revascularization (TLR), TVR, and TLF were 5.7%, 7.4%, and 11.4%, respectively. Conclusions: The safety and effectiveness of the R-ZES through 3 years in this real-world all-comer study was consistent with previously reported all-comer trials. (J Interven Cardiol 2013;26:515-523)

Description

Research Data

Keywords

Terms of Use

This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service

Endorsement

Review

Supplemented By

Related Stories