Publication:

Variability in MRI vs. ultrasound measures of prostate volume and its impact on treatment recommendations for favorable-risk prostate cancer patients: a case series

Loading...
Thumbnail Image

Date

2014

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

BioMed Central
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Murciano-Goroff, Y. R., L. D. Wolfsberger, A. Parekh, F. M. Fennessy, K. Tuncali, P. F. Orio, T. R. Niedermayr, et al. 2014. “Variability in MRI vs. ultrasound measures of prostate volume and its impact on treatment recommendations for favorable-risk prostate cancer patients: a case series.” Radiation Oncology (London, England) 9 (1): 200. doi:10.1186/1748-717X-9-200. http://dx.doi.org/10.1186/1748-717X-9-200.

Abstract

Background: Prostate volume can affect whether patients qualify for brachytherapy (desired size ≥20 mL and ≤60 mL) and/or active surveillance (desired PSA density ≤0.15 for very low risk disease). This study examines variability in prostate volume measurements depending on imaging modality used (ultrasound versus MRI) and volume calculation technique (contouring versus ellipsoid) and quantifies the impact of this variability on treatment recommendations for men with favorable-risk prostate cancer. Methods: We examined 70 patients who presented consecutively for consideration of brachytherapy for favorable-risk prostate cancer who had volume estimates by three methods: contoured axial ultrasound slices, ultrasound ellipsoid (height × width × length × 0.523) calculation, and endorectal coil MRI (erMRI) ellipsoid calculation. Results: Average gland size by the contoured ultrasound, ellipsoid ultrasound, and erMRI methods were 33.99, 37.16, and 39.62 mLs, respectively. All pairwise comparisons between methods were statistically significant (all p < 0.015). Of the 66 patients who volumetrically qualified for brachytherapy on ellipsoid ultrasound measures, 22 (33.33%) did not qualify on ellipsoid erMRI or contoured ultrasound measures. 38 patients (54.28%) had PSA density ≤0.15 ng/dl as calculated using ellipsoid ultrasound volumes, compared to 34 (48.57%) and 38 patients (54.28%) using contoured ultrasound and ellipsoid erMRI volumes, respectively. Conclusions: The ultrasound ellipsoid and erMRI ellipsoid methods appeared to overestimate ultrasound contoured volume by an average of 9.34% and 16.57% respectively. 33.33% of those who qualified for brachytherapy based on ellipsoid ultrasound volume would be disqualified based on ultrasound contoured and/or erMRI ellipsoid volume. As treatment recommendations increasingly rely on estimates of prostate size, clinicians must consider method of volume estimation.

Description

Research Data

Keywords

Prostate volume, Favorable-risk prostate cancer, Brachytherapy, Active surveillance, MRI, Ultrasound

Terms of Use

This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service