Evaluation of the FACSPresto, a New Point of Care Device for the Enumeration of CD4% and Absolute CD4+ T Cell Counts in HIV Infection

Loading...
Thumbnail Image

Open/View Files

Date

2016

Journal Title

Journal ISSN

Volume Title

Publisher

Public Library of Science
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Makadzange, Azure Tariro, Carola Bogezi, Kathryn Boyd, Anesu Gumbo, Dorinda Mukura, Allen Matubu, and Chiratidzo Ellen Ndhlovu. 2016. “Evaluation of the FACSPresto, a New Point of Care Device for the Enumeration of CD4% and Absolute CD4+ T Cell Counts in HIV Infection.” PLoS ONE 11 (7): e0157546. doi:10.1371/journal.pone.0157546. http://dx.doi.org/10.1371/journal.pone.0157546.

Abstract

Introduction: Enumeration of CD4+ T lymphocytes is important for pre-ART disease staging and screening for opportunistic infections, however access to CD4 testing in resource limited settings is poor. Point of care (POC) technologies can facilitate improved access to CD4 testing. We evaluated the analytical performance of a novel POC device the FACSPresto compared to the FACSCalibur as a reference standard and to the PIMA, a POC device in widespread use in sub-Saharan Africa. Method Specimens were obtained from 253 HIV infected adults. Venous blood samples were analyzed on the FACSPresto and the FACSCalibur, in a subset of 41 samples additional analysis was done on the PIMA. Results: The absolute CD4 count results obtained on the FACSPresto were comparable to those on the FACSCalibur with low absolute (9.5cells/μl) and relative bias (3.2%). Bias in CD4% values was also low (1.06%) with a relative bias of 4.9%. The sensitivity was lower at a CD4 count threshold of ≤350cells/μl compared with ≤500cells/μl (84.9% vs. 92.8%) resulting in a high upward misclassification rate at low CD4 counts. Specificity at thresholds of ≤350cells/μl and ≤500cells/μl were 96.6% and 96.8% respectively. The PIMA had a high absolute (-68.6cells/μl) and relative bias (-10.5%) when compared with the FACSCalibur. At thresholds of ≤350cells/μl and ≤500cells/μl the sensitivity was 100% and 95.5% respectively; specificity was 85.7% and 84.2% respectively. The coefficients of repeatability were 4.13%, 5.29% and 9.8% respectively. Discussion The analytic performance of the FACSPresto against the reference standard was very good with better agreement and precision than the PIMA. The FACSPresto had comparable sensitivity at a threshold of 500 cells/μl and better specificity than the PIMA. However the FACSPresto showed reduced sensitivity at low CD4 count thresholds. Conclusion: The FACSPresto can be reliably used as a POC device for enumerating absolute CD4 count and CD4% values.

Description

Research Data

Keywords

Biology and Life Sciences, Cell Biology, Cellular Types, Animal Cells, Blood Cells, White Blood Cells, T Cells, Immune Cells, Immunology, Medicine and Health Sciences, Anatomy, Body Fluids, Blood, Physiology, Hematology, Infectious Diseases, Opportunistic Infections, Pediatrics, Pediatric Infections, Medicine and health sciences, Infectious diseases, Viral diseases, HIV infections, Spectrum Analysis Techniques, Spectrophotometry, Cytophotometry, Flow Cytometry, People and Places, Geographical Locations, Africa, Zimbabwe

Terms of Use

This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service

Endorsement

Review

Supplemented By

Related Stories