Publication: What is the Value of Incorporating Tear Osmolarity Measurement in Assessing Patient Response to Therapy in Dry Eye Disease?
Open/View Files
Date
2014
Published Version
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier BV
The Harvard community has made this article openly available. Please share how this access benefits you.
Citation
Amparo, Francisco, Yiping Jin, Pedram Hamrah, Debra A. Schaumberg, and Reza Dana. 2014. “What Is the Value of Incorporating Tear Osmolarity Measurement in Assessing Patient Response to Therapy in Dry Eye Disease?” American Journal of Ophthalmology 157 (1) (January): 69–77.e2. doi:10.1016/j.ajo.2013.07.019.
Research Data
Abstract
Purpose
To evaluate the correlation between changes in tear osmolarity, symptoms, and corneal fluorescein staining in patients with dry eye disease (DED).
Design
Retrospective, clinic-based cohort study.
Methods
In this single-institution study, we reviewed the charts of 186 patients with DED from whom we had data on tear osmolarity, symptoms, and corneal fluorescein staining from two separate visits. Main outcomes included the correlation of the changes between the two visits for tear osmolarity (TearLab® system), symptoms (Ocular Surface Disease Index© [OSDI]), and corneal fluorescein staining (modified Oxford scheme). For tear osmolarity and corneal fluorescein staining the scores from the eye with highest readings were analyzed. The correlations were repeated on subgroups based on proposed cutoffs for DED severity and on patients’ treatment.
Results
We found a modest, though statistically significant, correlation between changes in corneal fluorescein staining and symptoms of DED (R = .31; P < .001). However, there was no correlation between the recorded change in tear osmolarity and symptoms (R = −.091; P = .38) or between changes in tear osmolarity and corneal fluorescein staining (R = −.02; P = .80). This lack of correlation was consistent in all the subgroups studied. A multivariate analysis revealed that changes in corneal fluorescein staining had predictive value on symptom changes, while tear osmolarity changes did not.
Conclusions
Changes in tear osmolarity do not correlate significantly with changes in patient symptoms or corneal fluorescein staining in dry eye disease.