Reduced exposure using asymmetric cone beam processing for wide area detector cardiac CT

Thumbnail Image

Open/View Files

Date

2011

Journal Title

Journal ISSN

Volume Title

Publisher

Springer Nature
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Bedayat, Arash, Frank J. Rybicki, Kanako Kumamaru, Sara L. Powers, Jason Signorelli, Michael L. Steigner, Chloe Steveson, et al. 2011. “Reduced Exposure Using Asymmetric Cone Beam Processing for Wide Area Detector Cardiac CT.” The International Journal of Cardiovascular Imaging 28 (2) (February 19): 381–388. doi:10.1007/s10554-011-9814-5.

Research Data

Abstract

The purpose of this study was to estimate dose reduction after implementation of asymmetrical cone beam processing using exposure differences measured in a water phantom and a small cohort of clinical coronary CTA patients. Two separate 320 × 0.5 mm detector row scans of a water phantom used identical cardiac acquisition parameters before and after software modifications from symmetric to asymmetric cone beam acquisition and processing. Exposure was measured at the phantom surface with Optically Stimulated Luminescence (OSL) dosimeters at 12 equally spaced angular locations. Mean HU and standard deviation (SD) for both approaches were compared using ROI measurements obtained at the center plus four peripheral locations in the water phantom. To assess image quality, mean HU and standard deviation (SD) for both approaches were compared using ROI measurements obtained at five points within the water phantom. Retrospective evaluation of 64 patients (37 symmetric; 27 asymmetric acquisition) included clinical data, scanning parameters, quantitative plus qualitative image assessment, and estimated radiation dose. In the water phantom, the asymmetric cone beam processing reduces exposure by approximately 20% with no change in image quality. The clinical coronary CTA patient groups had comparable demographics. The estimated dose reduction after implementation of the asymmetric approach was roughly 24% with no significant difference between the symmetric and asymmetric approach with respect to objective measures of image quality or subjective assessment using a four point scale. When compared to a symmetric approach, the decreased exposure, subsequent lower patient radiation dose, and similar image quality from asymmetric cone beam processing supports its routine clinical use.

Description

Keywords

Coronary imaging, Computed tomography, Angiography, Image reconstruction, Radiation dose