Publication: The effect of subcutaneous fat on electrical impedance myography when using a handheld electrode array: The case for measuring reactance
Open/View Files
Date
Published Version
Journal Title
Journal ISSN
Volume Title
Publisher
Citation
Abstract
Objective
Recent developments in electrical impedance myography (EIM) have led to the use of handheld electrode arrays (HEA) for data acquisition. Although preferable for several reasons, this approach tends to be more affected by subcutaneous fat (SF) than the original approach in which the impedance-measuring electrodes are widely spaced. In this study, we seek to identify the EIM parameter least impacted by subcutaneous fat (SF) when using an HEA.
Methods
18 normal subjects underwent 50 kHz EIM and ultrasound of the medial gastrocnemius muscles on the dominant side. Coefficients of determination (R2 values) were calculated for each of the three major EIM variables (reactance, resistance, and phase) and SF thickness.
Results
For both resistance and phase, a strong relationship to SF thickness was observed (R2 = 0.64 and R2 = 0.70, respectively, p < 0.001 for both). In contrast, for reactance, the relationship was non-significant, with R2 = 0.07, p = 0.30.
Conclusions
Unlike resistance and phase, both of which are highly impacted by SF thickness, the reactance shows no significant relationship.
Significance
Future clinical studies employing HEA’s to perform EIM should evaluate alterations in reactance over those in resistance and phase.