Publication: Titanium Coating of the Boston Keratoprosthesis
Open/View Files
Date
2016
Published Version
Journal Title
Journal ISSN
Volume Title
Publisher
The Association for Research in Vision and Ophthalmology
The Harvard community has made this article openly available. Please share how this access benefits you.
Citation
Salvador-Culla, Borja, Kyung Jae Jeong, Paraskevi Evi Kolovou, Homer H. Chiang, James Chodosh, Claes H. Dohlman, and Daniel S. Kohane. 2016. “Titanium Coating of the Boston Keratoprosthesis.” Translational Vision Science & Technology 5 (2): 17. doi:10.1167/tvst.5.2.17. http://dx.doi.org/10.1167/tvst.5.2.17.
Research Data
Abstract
Purpose We tested the feasibility of using titanium to enhance adhesion of the Boston Keratoprosthesis (B-KPro), ultimately to decrease the risk of implant-associated complications. Methods: Cylindrical rods were made of poly(methyl methacrylate) (PMMA), PMMA coated with titanium dioxide (TiO2) over a layer of polydopamine (PMMATiO2), smooth (Ti) and sandblasted (TiSB) titanium, and titanium treated with oxygen plasma (Tiox and TiSBox). Topography and surface chemistry were analyzed by scanning electron microscopy (SEM), atomic force microscopy (AFM), and X-ray photoelectron spectroscopy (XPS). Adhesion force between rods and porcine corneas was measured ex vivo. Titanium sleeves, smooth and sandblasted, were inserted around the stem of the B-KPro and implanted in rabbits. Tissue adhesion to the stem was assessed and compared to an unmodified B-Kpro after 1 month. Results: X-ray photoelectron spectroscopy demonstrated successful deposition of TiO2 on polydopamine-coated PMMA. Oxygen plasma treatment did not change the XPS spectra of titanium rods (Ti and TiSB), although it increased their hydrophilicity. The materials did not show cell toxicity. After 14 days of incubation, PMMATiO2, smooth titanium treated with oxygen plasma (Tiox), and sandblasted titanium rods (TiSB, TiSBox) showed significantly higher adhesion forces than PMMA ex vivo. In vivo, the use of a TiSB sleeve around the stem of the B-KPro induced a significant increase in tissue adhesion compared to a Ti sleeve or bare PMMA. Conclusions: Sandblasted titanium sleeves greatly enhanced adherence of the B-KPro to the rabbit cornea. This approach may improve adhesion with the donor cornea in humans as well. Translational Relevance This approach may improve adhesion with donor corneas in humans.
Description
Other Available Sources
Keywords
keratoprosthesis, Boston Keratoprosthesis, artificial cornea, titanium coating, titanium oxide, titanium skirt, biointegration, corneal adhesion, titanium adhesion
Terms of Use
This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service