MR-based motion correction for cardiac PET parametric imaging: a simulation study

Thumbnail Image

Open/View Files

Date

2018

Journal Title

Journal ISSN

Volume Title

Publisher

Springer International Publishing
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Guo, Rong, Yoann Petibon, Yixin Ma, Georges El Fakhri, Kui Ying, and Jinsong Ouyang. 2018. “MR-based motion correction for cardiac PET parametric imaging: a simulation study.” EJNMMI Physics 5 (1): 3. doi:10.1186/s40658-017-0200-9. http://dx.doi.org/10.1186/s40658-017-0200-9.

Research Data

Abstract

Background: Both cardiac and respiratory motions bias the kinetic parameters measured by dynamic PET. The aim of this study was to perform a realistic positron emission tomography-magnetic resonance (PET-MR) simulation study using 4D XCAT to evaluate the impact of MR-based motion correction on the estimation of PET myocardial kinetic parameters using PET-MR. Dynamic activity distributions were obtained based on a one-tissue compartment model with realistic kinetic parameters and an arterial input function. Realistic proton density/T1/T2 values were also defined for the MRI simulation. Two types of motion patterns, cardiac motion only (CM) and both cardiac and respiratory motions (CRM), were generated. PET sinograms were obtained by the projection of the activity distributions. PET image for each time frame was obtained using static (ST), gated (GA), non-motion-corrected (NMC), and motion-corrected (MC) methods. Voxel-wise unweighted least squares fitting of the dynamic PET data was then performed to obtain K1 values for each study. For each study, the mean and standard deviation of K1 values were computed for four regions of interest in the myocardium across 25 noise realizations. Results: Both cardiac and respiratory motions introduce blurring in the PET parametric images if the motion is not corrected. Conventional cardiac gating is limited by high noise level on parametric images. Dual cardiac and respiratory gating further increases the noise level. In contrast to GA, the MR-based MC method reduces motion blurring in parametric images without increasing noise level. It also improves the myocardial defect delineation as compared to NMC method. Finally, the MR-based MC method yields lower bias and variance in K1 values than NMC and GA, respectively. The reductions of K1 bias by MR-based MC are 7.7, 5.1, 15.7, and 29.9% in four selected 0.18-mL myocardial regions of interest, respectively, as compared to NMC for CRM. MR-based MC yields 85.9, 75.3, 71.8, and 95.2% less K1 standard deviation in the four regions, respectively, as compared to GA for CRM. Conclusions: This simulation study suggests that the MR-based motion-correction method using PET-MR greatly reduces motion blurring on parametric images and yields less K1 bias without increasing noise level.

Description

Keywords

MR-based PET motion correction, Cardiac PET parametric imaging, PET-MR, Myocardial perfusion