Publication: Measurement of gantry rotation time in modern ct
Open/View Files
Date
2014
Published Version
Journal Title
Journal ISSN
Volume Title
Publisher
John Wiley and Sons Inc.
The Harvard community has made this article openly available. Please share how this access benefits you.
Citation
Fukuda, Atsushi, Pei‐Jan P. Lin, Kosuke Matsubara, and Tosiaki Miyati. 2014. “Measurement of gantry rotation time in modern ct.” Journal of Applied Clinical Medical Physics 15 (1): 303-308. doi:10.1120/jacmp.v15i1.4517. http://dx.doi.org/10.1120/jacmp.v15i1.4517.
Research Data
Abstract
The purpose of this study was to develop and evaluate a noninvasive method to assess rotation time in modern commercial computed tomography (CT) systems. The rotation time was measured at a selected nominal rotation time (400 ms) utilizing two types of solid‐state detectors: the RTI's CT Dose Profiler (CTDP) and Unfors’ Xi (Xi) probes. Either CTDP or Xi was positioned on the inner cover of the gantry and a sheet of lead (1 mm thick) placed on top of the detector. Since a pair of two successive peaks is used to determine the gantry rotation time, by necessity the helical scan must be employed. Upon completion of the data acquisition, these peak times were determined with the dedicated software to obtain rotation time. The average rotation time obtained with CTDP and Xi operated under the dedicated software was found to be 400.6 and 400.5 ms, respectively. The detector for this measurement need not be specifically designed for CT dosimetry. The measurements of CT scanner rotation time can be accomplished with a radiation probe designed for the CT application or a conventional radiation probe designed for radiography and fluoroscopy applications. It is also noteworthy to point out that the measurement results are in good agreement between the two radiation detector systems. Finally, clinical medical physicists should be aware of the accuracy and precision of gantry rotation time, and take into consideration for QA where and when applicable. PACS number: 87.57.Q‐
Description
Other Available Sources
Keywords
Medical Imaging, gantry rotation time, solid‐state detector, CT, temporal resolution
Terms of Use
This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service