Determination of depth and field size dependence of multileaf collimator transmission in intensity‐modulated radiation therapy beams

Thumbnail Image

Date

2007

Published Version

Journal Title

Journal ISSN

Volume Title

Publisher

John Wiley and Sons Inc.
The Harvard community has made this article openly available. Please share how this access benefits you.

Research Projects

Organizational Units

Journal Issue

Citation

Zygmanski, Piotr, Florin Rosca, Dnyanesh Kadam, Friedlieb Lorenz, Adrian Nalichowski, Laurence Court, and Lee Chin. 2007. “Determination of depth and field size dependence of multileaf collimator transmission in intensity‐modulated radiation therapy beams.” Journal of Applied Clinical Medical Physics 8 (4): 76-95. doi:10.1120/jacmp.v8i4.2693. http://dx.doi.org/10.1120/jacmp.v8i4.2693.

Research Data

Abstract

Intensity‐modulated radiation therapy (IMRT) plans for the treatment of large and complex volumes may contain a relatively large contribution from multileaf collimator (MLC) transmission. In such cases, comprehensive characterization of direct and scatter MLC transmission is important. We designed a set of tests (open beam, closed static MLC, and dynamic MLC gap) to determine dosimetric MLC properties as a function of field size and depth at the central axis. We developed a generalized model of MLC transmission to account for direct MLC transmission, MLC scatter, beam hardening, and leaf‐end transmission (dosimetric gap). The model is consistent with the beam model used in IMRT optimization. We tested the model for extreme asymmetric fields relevant for large targets and for split IMRT fields. We applied our MLC scatter estimation formula to clinically relevant cases and showed that MLC scatter is contributing an undesired background dose. This contribution is relatively large, especially in low‐dose regions. (For instance, a uniform extra dose may dramatically increase normal‐lung toxicity in thorax treatment.) For complex IMRT of large‐volume targets, we found direct MLC transmission dose to be as high as 30%, and MLC scatter, up to 10% within the target volume for the selected cases. We identified that the dose discrepancies between the IMRT planning system [Eclipse (Varian Medical Systems, Palo Alto, CA)] and ionization chamber measurements (inside and outside of the field) are attributable to an inadequate model of MLC transmission in the planning system (constant‐value model). In the present study, we measured MLC transmission properties for Varian 6EX (6 MV) and 21EXs (6 and 10 MV) linear accelerators; however, the experimental method and theoretical model are more general. PACS number: 87.53.‐j

Description

Keywords

Radiation Oncology Physics, dynamic MLC, MLC transmission, dosimetry, IMRT, closed MLC

Terms of Use

This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service

Endorsement

Review

Supplemented By

Referenced By

Related Stories