Publication: Toward noninvasive monitoring of ongoing electrical activity of human uterus and fetal heart and brain
Open/View Files
Date
2017
Authors
Published Version
Journal Title
Journal ISSN
Volume Title
Publisher
Elsevier
The Harvard community has made this article openly available. Please share how this access benefits you.
Citation
Lew, S., M.S. Hämäläinen, and Y. Okada. 2017. “Toward noninvasive monitoring of ongoing electrical activity of human uterus and fetal heart and brain.” Clinical Neurophysiology 128 (12): 2470-2481. doi:10.1016/j.clinph.2017.08.026. http://dx.doi.org/10.1016/j.clinph.2017.08.026.
Research Data
Abstract
Objective: To evaluate whether a full-coverage fetal-maternal scanner can noninvasively monitor ongoing electrophysiological activity of maternal and fetal organs. Methods: A simulation study was carried out for a scanner with an array of magnetic field sensors placed all around the torso from the chest to the hip within a horizontal magnetic shielding enclosure. The magnetic fields from internal organs and an external noise source were computed for a pregnant woman with a 35-week old fetus. Signal processing methods were used to reject the external and internal interferences, to visualize uterine activity, and to detect activity of fetal heart and brain. Results: External interference was reduced by a factor of 1000, sufficient for detecting signals from internal organs when combined with passive and active shielding. The scanner rejects internal interferences better than partial-coverage arrays. It can be used to estimate currents around the uterus. It clearly detects spontaneous activity from the fetal heart and brain without averaging and weaker evoked brain activity at all fetal head positions after averaging. Conclusion: The simulated device will be able to monitor the ongoing activity of the fetal and maternal organs. Significance This type of scanner may become a novel tool in fetal medicine.
Description
Other Available Sources
Keywords
Prenatal medicine, Magnetocardiography (MCG), Electrocardiography (ECG), Electrohysterography (EHG), Magnetoencephalography (MEG), Electroencephalography (EEG)
Terms of Use
This article is made available under the terms and conditions applicable to Other Posted Material (LAA), as set forth at Terms of Service